Tel:
+49 (0)241 95 163 153
Fax:
+49 (0)241 95 163 155
E-Mail:
orders@anticorps-enligne.fr

Cotinine Kit ELISA

Reactivité: Souris, Rat Colorimetric Competition ELISA Urine
N° du produit ABIN1721161
  • Antigène Voir toutes Cotinine Kits ELISA
    Cotinine
    Reactivité
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Souris, Rat
    Méthode de détection
    Colorimetric
    Type de méthode
    Competition ELISA
    Seuil minimal de détection
    1 ng/mL
    Application
    ELISA
    Fonction
    The Cotinine kit is a solid phase competitive ELISA. The samples and Cotinine enzyme conjugate are added to the wells coated with anti-Cotinine antibody. Cotinine in the samples competes with a Cotinine enzyme (HRP) conjugate for binding sites. Unbound Cotinine and Cotinine enzyme conjugate is washed off by washing step. Upon the addition of the substrate, the intensity of color is inversely proportional to the concentration of Cotinine in the samples. A standard curve is prepared relating color intensity to the concentration of the Cotinine.
    Type d'échantillon
    Urine
    Analytical Method
    Quantitative
    Specificité
    The specificity of this Cotinine ELISA was determined by generating inhibition curves for each of the compounds listed below the antisera cross-reactivity below:
    CompoundApprox. ng/mL equivalent to 100 ng Cotinine/mLCross-reactivity
    Cotinine100100
    Nicotinemore than 10.000less than 1
    Nicotinamidemore than 10.000less than 1
    Nicotinic Acidmore than 10.000less than 1
    Réactivité croisée (Details)
    Aliquots of a human urine matrix were spiked with the following compounds at a concentration of 50,000 ng/mL. None of these compounds gave values in the assay that were equal to or greater than the assay sensitivity level. Acetaminophen, Acetylsalicylic acid, Amphetamine, Aminopyrine, Ampicillin, Amobarbital, Ascorbic acid, Atropine , Barbital, Butabarbital, Caffeine , Cocaine, Carbamazepine, Codeine , Chloroquine, Chloropromazine, Carbromal , Desipramine, Dextromethorphan, Dextropropoxyphene , 5,5-Diphenylhydantoin, 10-11-Dihydrocarbamazepine, Diazepam, Ethosuximide, Estriol, Estrone, Estradiol, Ethotoin, Glutethimide , Hexobarbital, Ibuprofen, Imipramine, Lidocaine, LSD, Methadone , Methadone-primary metabolite, Methaqualone, Methamphetamine, Metharbital, Mephenytoin, Mephobarbital, Methyl PEMA, Ibuprofen, Imipramine, Lidocaine, LSD, Methadone , Methadone-primary metabolite, Methaqualone, Methamphetamine, Metharbital, Mephenytoin, Mephobarbital, Methyl PEMA, Methsuximide , 4-Methylprimidone, Morphine , Meperidine , Niacinamide, Norethindrone, N-Normethsuximide, Phenobarbital , Phensuximide, PEMA, Primidone, Phencyclidine, Pentobarbital, Phenothiazine, Phenylpropanolamine, Procaine, Quinine, Secobarbital, Tetracycline,Tetrahydrozoline
    Sensibilité
    Assay sensitivity based on the minimum Cotinine concentration required to produce a three standard deviation from assay Ao is 1 ng/mL.
    Attributs du produit
    • This Cotinine Direct ELISA Kit is to be used with Mouse/Rat urine or serum. This assay has not tested for all possible applications. Cutoff criteria are important in deciding the sample dilution.
    • Specimens to which sodium azide has been added affect the assay.
    Ingrédients
    1. Microwell coated with polyclonal Ab to Cotinine: 12 x 8 x 1
    2. Standard Set (ready to use): 0.5 mL
    3. Cotinine HRP Enzyme Conjugate (ready to use): 12 mL
    4. TMB Substrate (ready to use): 12 mL
    5. Stop Solution (ready to use): 12 mL
    Matériel non inclus
    1. Distilled or deionized water
    2. Precision pipettes
    3. Disposable pipette tips
    4. ELISA reader capable of reading absorbance at 450 nm
    5. Absorbance paper or paper towel
    6. Graph paper
    Featured
    Discover our best selling Cotinine Kit ELISA
  • Indications d'application
    The Calbiotech Cotinine Direct ELISA Kit is intended for the measurement of Cotinine in Mouse/Rat serum or urine.
    Volume d'échantillon
    100 μL
    Durée du test
    2 - 3 h
    Plaque
    Pre-coated
    Protocole
    All reagents must be brought to room temperature (18-26 °C) before use.
    • Pipette 10 µL of standards, controls and specimens into selected well in duplicate.
    • Add 100 µL of the Enzyme Conjugate to each well. Shake the plate, 10-30 s, to ensure proper mixing.
    • Incubate for 60 min at room temperature (18-26 °C) preferably in the dark.
    • Wash the wells 6 times with 300 µL distilled water using either a suitable plate washer or wash bottle taking care not to cross contaminate wells.
    • Invert wells and vigorously slap dry on absorbent paper to ensure all residual moisture is removed. This step is critical to ensure that residual enzyme conjugate, does not skew results. If using an automated system, ensure that the final aspiration on the wash cycle aspirates from either side of the well.
    • Add 100 µL of Substrate reagent to each well.
    • Incubate for 30 min at room temperature, preferably in the dark.
    • Add 100 µL of Stop Solution to each well. Shake the plate gently to mix the solution.
    • Read absorbance on ELISA Reader at 450 nm with in 15 min after adding the stopping solution.
    Calcul des résultats

    The standard curve is constructed as follows:

    1. Check Cotinine standard value on each standard vial.
    2. To construct the standard curve, plot the absorbance for Cotinine standards (vertical axis) versus Cotinine standard concentrations (horizontal axis) on a linear graph paper. Draw the best curve through the points.
    3. Read the absorbance for controls and each unknown sample from the curve. Record the value for each control or unknown sample.

    Précision du teste
    20 urine samples from non smokers were screened with this Cotinine ELISA method. All 20 samples screened negative with the ELISA method. 15 samples from smokers which contained various amounts of Cotinine were screened with This Cotinine Direct ELISA Kit. All 15 samples showed a presence of Cotinine at a level greater than 500 ng/mL. Three urine samples submitted by individuals exposed to passive inhalation for over 30 days all showed levels of 5 to 10 ng/mL of Cotinine when extrapolated of a dose response curve.
    Restrictions
    For Research Use only
  • Validation #029687 (ELISA)
    'Independent Validation' signe
    by
    Affina Biotechnologies, Inc
    No.
    #029687
    Date
    24.04.2014
    Antigène
    Numéro du lot
    COT4179
    Application validée
    ELISA
    Contrôle positif
    Female rat serum spiked with cotinine
    Contrôle négative
    Unspiked female rat serum
    Conclusion
    Signal was detected in positive control sample and not in negative control sample.
    'Independent Validation' signe
    Validation Images
    Protocole
    Anticorps primaire
    • Antigen: Cotinine
    • Catalog number: ABIN1721161
    • Lot number: COT4179
    Anticorps secondaire
    Full Protocol
    • 10 µL of standard and samples were added to the 96-well strip plates provided in the kit and mixed with working solution of enzyme conjugate. All samples and standards were assayed in duplicate.
    • The microplate was covered and incubated at RT for 60 min.
    • Content of the wells was discarded and wells were washed 6 times with 250 µL of water.
    • 100 µL of substrate was added to each well. The plate was covered and incubated at RT for 30 min in the dark.
    • 100 µL of the Stop Solution was added per well.
    • The entire sample was transferred into a 96-well plate (Nunc, Maxisorp)
    • The optical density (OD value) of each well was read immediately using a microplate reader set to 450 nm.
    • The duplicate readings for each sample were averaged and the average zero standard optical density subtracted. The corrected average-value was tabulated as Average Absorbance. A standard curve was generated by plotting the mean OD value for each standard on the x-axis against the concentration on the Y-axis using Kaleidagraph. The concentration of samples was calculated using the CurveExpert v 1.4 logistic fit (Concentration = 3.52/(1-e-6.8*OD450).
    Notes
    Nothing noted.
  • Précaution d'utilisation
    Potential biohazardous materials: The calibrator and controls contain human source components which have been tested and found non-reactive for hepatitis B surface antigen as well as HIV antibody with FDA licensed reagents. However, as there is no test method that can offer complete assurance that HIV, Hepatitis B virus or other infectious agents are absent, these reagents should be handled at the Biosafety Level 2, recommended in the Centers for Disease Control/National Institutes of Health manual,
    Conseil sur la manipulation
    Keep microwells sealed in a dry bag with desiccants
    The reagents are stable until expiration of the kit.
    Do not expose test reagents to heat, sun, or strong light.
    Stock
    4 °C
  • Antigène Voir toutes Cotinine Kits ELISA
    Cotinine
    Abstract
    Cotinine Produits
    Classe de substances
    Chemical
    Sujet
    Exposure to tobacco smoke can be detected by measuring nicotine and its metabolites. Nicotine has a short half life and is not used as a marker for tobacco smoke exposure. Cotinine due to its longer half life has been used in research as a reliable marker for smoking status and smoking cessation studies. The Calbiotech Cotinine Direct ELISA Kit is designed for the detection Cotinine in rat serum and urine. It can also be adapted for other fluids.
    Numéro CAS
    486-56-6
Vous êtes ici:
Support technique