CTH Kit ELISA

CTH Reactivité: Souris Colorimetric Sandwich ELISA 0.31 ng/mL - 20 ng/mL Plasma, Serum
N° du produit ABIN424908
  • Antigène Voir toutes CTH Kits ELISA
    CTH (Cystathionase (Cystathionine gamma-Lyase) (CTH))
    Reactivité
    • 5
    • 4
    • 2
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Souris
    Méthode de détection
    Colorimetric
    Type de méthode
    Sandwich ELISA
    Gamme de detection
    0.31 ng/mL - 20 ng/mL
    Seuil minimal de détection
    0.31 ng/mL
    Application
    ELISA
    Type d'échantillon
    Plasma, Serum
    Sensibilité
    0.112 ng/mL
    Ingrédients
    • Pre-coated, ready to use 96-well strip plate, flat buttom
    • Plate sealer for 96 wells
    • Reference Standard
    • Standard Diluent
    • Detection Reagent A
    • Detection Reagent B
    • Assay Diluent A
    • Assay Diluent B
    • Reagent Diluent (if Detection Reagent is lyophilized)
    • TMB Substrate
    • Stop Solution
    • Wash Buffer (30 x concentrate)
    • Instruction manual
    Matériel non inclus
    • Microplate reader with 450nm filter.
    • Precision single or multi-channel pipettes and disposable tips.
    • Eppendorf Tubes for diluting samples.
    • Deionized or distilled water.
    • Absorbent paper for blotting the microtiter plate.
    • Container for Wash Solution
    Featured
    Discover our best selling CTH Kit ELISA
    Top Product
    Discover our top product CTH Kit ELISA
  • Indications d'application
    • Limited by the current condition and scientific technology, we cannot completely conduct the comprehensive identification and analysis on the raw material provided by suppliers. So there might be some qualitative and technical risks to use the kit.
    • The final experimental results will be closely related to validity of the products, operation skills of the end users and the experimental environments. Please make sure that sufficient samples are available.
    • Kits from different batches may be a little different in detection range, sensitivity and color developing time.
    • Do not mix or substitute reagents from one kit lot to another. Use only the reagents supplied by manufacturer.
    • Protect all reagents from strong light during storage and incubation. All the bottle caps of reagents should be covered tightly to prevent the evaporation and contamination of microorganism.
    • There may be some foggy substance in the wells when the plate is opened at the first time. It will not have any effect on the final assay results. Do not remove microtiter plate from the storage bag until needed.
    • Wrong operations during the reagents preparation and loading, as well as incorrect parameter setting for the plate reader may lead to incorrect results. A microplate plate reader with a bandwidth of 10nm or less and an optical density range of 0-3 O.D. or greater at 450 ± 10nm wavelength is acceptable for use in absorbance measurement. Please read the instruction carefully and adjust the instrument prior to the experiment.
    • Even the same operator might get different results in two separate experiments. In order to get better reproducible results, the operation of every step in the assay should be controlled. Furthermore, a preliminary experiment before assay for each batch is recommended.
    • Each kit has been strictly passed Q.C test. However, results from end users might be inconsistent with our in-house data due to some unexpected transportation conditions or different lab equipments. Intra-assay variance among kits from different batches might arise from above factors, too.
    • Kits from different manufacturers for the same item might produce different results, since we have not compared our products with other manufacturers.
    Commentaires

    Information on standard material:
    The standard might be recombinant protein or natural protein, that will depend on the specific kit. Moreover, the expression system is E.coli or yeast or mammal cell. There is 0.05% proclin 300 in the standard as preservative.

    Information on reagents:
    The stop solution used in the kit is sulfuric acid with concentration of 1 mol/L. And the wash solution is TBS. The standard diluent contains 0.02 % sodium azide, assay diluent A and assay diluent B contain 0.01% sodium azide. Some kits can contain is BSA in them.

    Information on antibodies:
    The provided antibodies and their host vary in different kits.

    Volume d'échantillon
    100 μL
    Plaque
    Pre-coated
    Préparation des réactifs
    1. Bring all kit components and samples to room temperature (18-25 °C) before use. If the kit will not be used up in one time, please only take out strips and reagents for present experiment, and leave the remaining strips and reagents in required condition.
    2. Standard - Reconstitute the Standard with 1.0 mL of Standard Diluent, keep for 10 minutes at room temperature, shake gently (not to foam). The concentration of the standard in the stock solution is 80 ng/mL. Firstly dilute the stock solution to 20 ng/mL and the diluted standard serves as the highest standard (20 ng/mL). Then prepare 7 tubes containing 0.5 mL Standard Diluent and use the diluted standard to produce a double dilution series. Mix each tube thoroughly before the next transfer. Set up 7 points of diluted standard such as 20 ng/mL, 10 ng/mL, 5 ng/mL, 2.5 ng/mL, 1.25 ng/mL, 0.625 ng/mL, 0.312 ng/mL, and the last microcentrifuge tube with Standard Diluent is the blank as 0 ng/mL.
    3. Detection Reagent A and Detection Reagent B - If lyophilized reconstitute the Detection Reagent A with 150μL of Reagent Diluent, keep for 10 minutes at room temperature, shake gently (not to foam). Briefly spin or centrifuge the stock Detection A and Detection B before use. Dilute them to the working concentration 100-fold with Assay Diluent A and B, respectively.
    4. Wash Solution - Dilute 20 mL of Wash Solution concentrate (30x) with 580 mL of deionized or distilled water to prepare 600 mL of Wash Solution (1x).
    5. TMB substrate - Aspirate the needed dosage of the solution with sterilized tips and do not dump the residual solution into the vial again.

    Note:

    1. Making serial dilution in the wells directly is not permitted.
    2. Prepare standards within 15 minutes before assay. Please do not dissolve the reagents at 37 °C directly.
    3. Please carefully reconstitute Standards or working Detection Reagent A and B according to the instruction, and avoid foaming and mix gently until the crystals are completely dissolved. To minimize imprecision caused by pipetting, use small volumes and ensure that pipettors are calibrated. It is recommended to suck more than 10μL for one pipetting.
    4. The reconstituted Standards, Detection Reagent A and Detection Reagent B can be used only once.
    5. If crystals have formed in the Wash Solution concentrate (30x), warm to room temperature and mix gently until the crystals are completely dissolved.
    6. Contaminated water or container for reagent preparation will influence the detection result.
    Prélèvement de l'échantillon
    Serum: Allow samples to clot for two hours at room temperature or overnight at 4°C before centrifugation for 20 minutes at approximately 1000 × g. Assay immediately or store samples in aliquot at -20°C or -80°C. Avoid repeated freeze/thaw cycles.

    Plasma: Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 15 minutes at 1000 × g within 30 minutes of collection. Remove plasma and assay immediately or store samples in aliquot at -20°C or -80°C. Avoid repeated freeze/thaw cycles.

    Biological Fluids: Centrifuge samples for 20 minutes at 1000 × g. Remove particulates and assay immediately or store samples in aliquot at -20 °C or -80 °C for later use. Avoid repeated freeze/thaw cycles.
    Calcul des résultats

    Average the duplicate readings for each standard, control, and samples and subtract the average zero standard optical density. Construct a standard curve by plotting the mean O.D. and concentration for each standard and draw a best fit curve through the points on the graph or create a standard curve on log-log graph paper with the target concentration on the y-axis and absorbance on the x-axis. Draw the best fit straight line through the standard points and it can be determined by regression analysis.Using some plot software, for instance, curve expert 1.30, is also recommended. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.

    Précision du teste
    Intra-assay Precision (precision within an assay): Three samples with low, medium and high levels of the target antigen were tested twenty times on one plate, respectively.
    Inter-assay Precision (precision between assays): Three samples with low, medium and high levels of the target antigen were tested on three different plates, eight replicates in each plate.
    CV (%) = SD/mean X 100
    • Intra-assay: CV less than 10 %
    • Inter-assay: CV less than 12 %
    Restrictions
    For Research Use only
  • Précaution d'utilisation
    The Stop Solution suggested for use with this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.
    Conseil sur la manipulation
    To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
    Stock
    4 °C/-20 °C
    Stockage commentaire
    • For unopened kit: All the reagents should be kept according to the labels on vials. The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20°C upon receipt while the others should be at 4°C.
    • For opened kit: When the kit is opened, the remaining reagents still need to be stored according to the above storage condition. Besides, please return the unused wells to the foil pouch containing the desiccant pack, and reseal along entire edge of zip-seal.
      Note: It is highly recommended to use the remaining reagents within 1 month provided this is within the expiration date of the kit.
    • For ELISA kit, 1 day storage at 37°C can be considered as 2 months at 4°C, which means 3 days at 37°C equaling 6 months at 4°C.
    Date de péremption
    6 months
  • Antigène Voir toutes CTH Kits ELISA
    CTH (Cystathionase (Cystathionine gamma-Lyase) (CTH))
    Autre désignation
    Cystathionine Gamma Lyase (CSE) (CTH Produits)
    Synonymes
    DYT17 Kit ELISA, DYT18 Kit ELISA, DYT9 Kit ELISA, EIG12 Kit ELISA, GLUT Kit ELISA, GLUT1 Kit ELISA, GLUT1DS Kit ELISA, HTLVR Kit ELISA, PED Kit ELISA, wu:fb48e03 Kit ELISA, wu:fb58d08 Kit ELISA, zgc:66001 Kit ELISA, zgc:85785 Kit ELISA, CTH Kit ELISA, 0610010I13Rik Kit ELISA, AI314617 Kit ELISA, BC019483 Kit ELISA, CSE Kit ELISA, Cys3 Kit ELISA, cysA Kit ELISA, SCQ11.03c Kit ELISA, DDBDRAFT_0190313 Kit ELISA, DDBDRAFT_0191318 Kit ELISA, DDB_0190313 Kit ELISA, DDB_0191318 Kit ELISA, H Kit ELISA, MGC75946 Kit ELISA, cystathionine gamma-lyase Kit ELISA, solute carrier family 2 member 1 Kit ELISA, cystathionase (cystathionine gamma-lyase) Kit ELISA, cystathionine gamma-lyase L homeolog Kit ELISA, CTH Kit ELISA, SLC2A1 Kit ELISA, cth Kit ELISA, Cth Kit ELISA, SCO3920 Kit ELISA, mlr1566 Kit ELISA, TA_RS00415 Kit ELISA, MAP_RS15640 Kit ELISA, SPO3344 Kit ELISA, CNN01730 Kit ELISA, Mmcs_1928 Kit ELISA, RHA1_RS31820 Kit ELISA, HNE_1212 Kit ELISA, MAV_3867 Kit ELISA, MSMEG_2394 Kit ELISA, Mvan_2142 Kit ELISA, SACE_6136 Kit ELISA, Mflv_4220 Kit ELISA, cysA Kit ELISA, cth.L Kit ELISA
    Pathways
    ER-Nucleus Signaling, L'effet Warburg
Vous êtes ici:
Support technique