FBLN3 Kit CLIA

FBLN3 Reactivité: Humain Chemiluminescent Sandwich ELISA 1.56 ng/mL - 100 ng/mL Plasma, Serum
N° du produit ABIN5654335
  • Antigène Voir toutes FBLN3 Kits CLIA
    FBLN3 (Fibulin 3 (FBLN3))
    Reactivité
    • 9
    • 4
    • 1
    Humain
    Méthode de détection
    Chemiluminescent
    Type de méthode
    Sandwich ELISA
    Gamme de detection
    1.56 ng/mL - 100 ng/mL
    Seuil minimal de détection
    1.56 ng/mL
    Application
    ELISA
    Type d'échantillon
    Plasma, Serum
    Analytical Method
    Quantitative
    Specificité
    This assay has high sensitivity and excellent specificity for detection of Fibulin 3 (FBLN3). No significant cross-reactivity or interference between Fibulin 3 (FBLN3) and analogues was observed.
    Sensibilité
    0.07 ng/mL
    Featured
    Discover our best selling FBLN3 Kit ELISA
    Top Product
    Discover our top product FBLN3 Kit ELISA
  • Commentaires

    The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

    Durée du test
    2 - 3 h
    Plaque
    Pre-coated
    Protocole
    The microplate provided in this kit has been pre-coated with an antibody specific to Fibulin 3 (FBLN3). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Fibulin 3 (FBLN3). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Fibulin 3 (FBLN3) level in the sample or standard.,
    Précision du teste
    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Fibulin 3 (FBLN3) were tested 20 times on one plate, respectively
    Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Fibulin 3 (FBLN3) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100
    Intra-Assay: CV<10%
    Inter-Assay: CV<12%
    Restrictions
    For Research Use only
  • Conseil sur la manipulation
    Do not allow to contact skin or eyes. Calibrators, controls and specimen samples should be assayed in duplicate. Once the procedure has been started, all steps should be completed without interruption.
    Stock
    4 °C,-20 °C
    Stockage commentaire
    -20°C. Bring all reagents to room temperature before beginning test. The kit may be stored at 4°C for immediate use within two days upon arrival. Reseal any unused strips with desiccant pack. Minimize freeze/thaw cycles.
    Date de péremption
    4-8 months
  • Antigène Voir toutes FBLN3 Kits CLIA
    FBLN3 (Fibulin 3 (FBLN3))
    Autre désignation
    Fibulin 3 (FBLN3 Produits)
    Synonymes
    DHRD Kit CLIA, DRAD Kit CLIA, FBLN3 Kit CLIA, FBNL Kit CLIA, FIBL-3 Kit CLIA, MLVT Kit CLIA, MTLV Kit CLIA, S1-5 Kit CLIA, T16 Kit CLIA, EGF containing fibulin like extracellular matrix protein 1 Kit CLIA, EGF-containing fibulin-like extracellular matrix protein 1 Kit CLIA, epidermal growth factor-containing fibulin-like extracellular matrix protein 1 Kit CLIA, EFEMP1 Kit CLIA, Efemp1 Kit CLIA
    Sujet

    Gene Name: Fibulin 3

    Gene Aliases: EFEMP1, DHRD, DRAD, FBNL, MLVT, MTLV, S1-5, EGF-Containing Fibulin-Like Extracellular Matrix Protein 1, Fibrillin-Like, Extracellular protein S1-5

    ID gène
    2202
    UniProt
    Q12805
    Pathways
    EGFR Signaling Pathway