Vip Kit ELISA
-
- Antigène Voir toutes Vip Kits ELISA
- Vip (Vasoactive Intestinal Peptide (Vip))
-
Reactivité
- Rat
- Méthode de détection
- Colorimetric
- Type de méthode
- Sandwich ELISA
- Application
- ELISA
- Fonction
- This immunoassay kit allows for the specific measurement of rat Vasoactive intestinal peptide ,VIP concentrations in cell culture supernates, serum, and plasma.
- Type d'échantillon
- Cell Culture Supernatant, Serum, Plasma
- Analytical Method
- Quantitative
- Specificité
- This assay recognizes recombinant and natural rat Vasoactive Intestinal Peptide.
- Réactivité croisée (Details)
- No significant cross-reactivity or interference was observed.
- Sensibilité
- The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as the lowest detectable concentration that could be differentiated from zero.
- Attributs du produit
- Rattus norvegicus,Rat,VIP peptides,Vip
- Ingrédients
- Reagent (Quantity ): Assay plate (1), Standard (2), Sample Diluent (1 × 20ml), Assay Diluent A (1x10ml), Assay Diluent B (1x10ml), Detection Reagent A (1 × 120μl), Detection Reagent B (1 × 120μl), Wash Buffer (25 x concentrate) (1 × 30ml), Substrate (1x10ml), Stop Solution (1 x 10ml)
- Featured
- Discover our best selling Vip Kit ELISA
- Top Product
- Discover our top product Vip Kit ELISA
-
-
- Volume d'échantillon
- 100 μL
- Plaque
- Pre-coated
- Protocole
- This assay employs the quantitative sandwich enzyme immunoassay technique. A monoclonal antibody specific for Vasoactive Intestinal Peptide has been pre-coated onto a microplate. Standards and samples are pipetted into the wells and any Vasoactive Intestinal Peptide present is bound by the immobilized antibody. An enzyme-linked monoclonal antibody specific for Vasoactive Intestinal Peptide is added to the wells. Following a wash to remove any unbound antibody-enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Vasoactive Intestinal Peptide bound in the initial step. The color 2 development is stopped and the intensity of the color is measured.
- Préparation des réactifs
-
Bring all reagents to room temperature before use. Wash Buffer - If crystals have formed in the concentrate, warm to room temperature and mix gently until the crystals have completely dissolved. Dilute 30 mL of Wash Buffer Concentrate into deionized or distilled water to prepare 750 mL of Wash Buffer. Standard - Reconstitute the Standard with 1.0 ml of Sample Diluent. This reconstitution produces a stock solution of 80 ng/mL. Allow the standard to sit for about 10 minutes with gentle agitation prior to making serial dilutions (Making serial dilution in the wells directly is not permitted). Please firstly dilute the stock solution to 40 ng/mL and the diluted standard serves as the high standard ( 40 ng/mL). The Sample Diluent serves as the zero standard (0 ng/mL). ng/mL 80 40 20 10 5 2.5 1.25 0.625 0 Detection Reagent A and B - Dilute to the working concentration using Assay Diluent A or B (1:100), respectively.
4. - Prélèvement de l'échantillon
- Tissue homogenates - The preparation of tissue homogenates will vary depending upon tissue type. For this assay, tissue was rinsed with 1X PBS to remove excess blood, homogenized in 5~10 mL of 1X PBS and stored overnight at ≤ -20 °C. After two freeze-thaw cycles were performed to break the cell membranes, the homogenates were centrifuged for 5 minutes at 5000 x g. Remove the supernate and assay immediately or aliquot and store at ≤ -20 °C. Other biological fluids - Remove particulates by centrifugation and assay immediately or aliquot and store samples at -20 or -80 . Avoid repeated freeze-thaw cycles. Note: Tissue homogenates and other biological fluids to be used within 7 days may be stored at 2-8, otherwise samples must stored at -20 ( 1 month) or -80 ( 2 months) 3 to avoid loss of bioactivity and contamination. Avoid freeze-thaw cycles. When performing the assay slowly bring samples to room temperature.
- Procédure de l'essai
-
Allow all reagents to reach room temperature (Please do not dissolve the reagents at 37 directly.). All the reagents should be mixed thoroughly by gently swirling before pipetting. Avoid foaming. Keep appropriate numbers of strips for 1 experiment and remove extra strips from microtiter plate. Removed strips should be resealed and stored at 4 until the kits expiry date. Prepare all reagents, working standards and samples as directed in the previous sections. Please predict the concentration before assaying. If values for these are not within the range of the standard curve, users must determine the optimal sample dilutions for their particular experiments.
1. Add 100 of Standard, Blank, or Sample per well. Cover with the Plate sealer. Incubate for two hours at 37 .
2. Remove the liquid of each well, don ’ t wash.
3. Add 100 μ l of Detection Reagent A working solution to each well. Cover with the Plate sealer. Incubate for 1 hour at 37 . Detection Reagent A working solution may appear cloudy. Warm to room temperature and mix gently until solution appears uniform.
4. Aspirate each well and wash, repeating the process three times for a total of three washes. Wash by filling each well with Wash Buffer (approximately 400 μ l) using a squirt bottle, multi-channel pipette, manifold dispenser or autowasher. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Buffer by aspirating or decanting. Invert the plate and blot it against clean paper towels.
5. Add 100 μ l of Detection Reagent B working solution to each well. Cover with a new Plate sealer. Incubate for one hour at 37 .
6. Repeat the aspiration/wash process for five times as conducted in step
4. 7. Add 90 μ l of Substrate Solution to each well. Cover with a new Plate sealer. Incubate for 15 - 30 minutes at 37 . Protect from light.
8. Add 50 μ l of Stop Solution to each well. If color change does not appear uniform, gently tap the plate to ensure thorough mixing.
9. Determine the optical density of each well at once, using a microplate reader set to 450 nm.
Important Note:1. Absorbance is a function of the incubation time. Therefore, prior to starting the assay it is recommended that all reagents should be freshly prepared prior to use and all required strip-wells are secured in the microtiter frame. This will ensure equal elapsed time for each pipetting step, without interruption.
2. Please carefully reconstitute Standards or working Detection Reagent A and B according to the instruction, and avoid foaming and mix gently until the crystals have completely dissolved. The reconstituted Standards can be used only once. This assay requires pipetting of small volumes. To minimize imprecision caused by pipetting, ensure that pipettors are calibrated. It is recommended to suck more than 10 μ l for once pipetting. 5
3. To ensure accurate results, proper adhesion of plate sealers during incubation steps is necessary. Do not allow wells to sit uncovered for extended periods between incubation steps. Once reagents have been added to the well strips, DO NOT let the strips DRY at any time during the assay.
4. For each step in the procedure, total dispensing time for addition of reagents to the assay plate should not exceed 10 minutes.
5. To avoid cross-contamination, change pipette tips between additions of each standard level, between sample additions, and between reagent additions. Also, use separate reservoirs for each reagent.
6. The wash procedure is critical. Insufficient washing will result in poor precision and falsely elevated absorbance readings.
7. Duplication of all standards and specimens, although not required, is recommended.
8. Substrate Solution is easily contaminated. Please protect it from light. - Calcul des résultats
-
Average the duplicate readings for each standard, control, and samples and subtract the average zero standard optical density. Create a standard curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. As an alternative, construct a standard curve by plotting the mean absorbance for each standard on the x-axis against the concentration on the y-axis and draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the PI3K concentrations versus the log of the O.D. and the best fit line can be determined by regression analysis. It is recommended to use some related software to do this calculation, such as curve expert 13.0. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.
- Restrictions
- For Research Use only
-
- Conseil sur la manipulation
-
1. The kit should not be used beyond the expiration date on the kit label.
2. Do not mix or substitute reagents with those from other lots or sources.
3. If samples generate values higher than the highest standard, further dilute the samples and repeat the assay. Any variation in standard diluent, operator, pipetting technique, washing technique,incubation time or temperature, and kit age can cause variation in binding.
4. This assay is designed to eliminate interference by soluble receptors, ligands, binding proteins, and other factors present in biological samples. Until all factors have been tested in the Immunoassay, the possibility of interference cannot be excluded.
5. Limited by the current condition and scientific technology, we can't completely conduct the comprehensive identification and analysis on the raw material provided by suppliers. So there might be some qualitative and technical risks to use the kit. - Stock
- 4 °C/-20 °C
- Stockage commentaire
- The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20 °C upon being received. The other reagents can be stored at 4 °C.
-
- Antigène Voir toutes Vip Kits ELISA
- Vip (Vasoactive Intestinal Peptide (Vip))
- Autre désignation
- Vip (Vip Produits)
- Synonymes
- PHM27 Kit ELISA, vip1 Kit ELISA, zgc:171463 Kit ELISA, VIP Kit ELISA, vip-A Kit ELISA, PHI Kit ELISA, vasoactive intestinal peptide Kit ELISA, vasoactive intestinal peptide L homeolog Kit ELISA, vasoactive intestinal polypeptide Kit ELISA, VIP Kit ELISA, Vip Kit ELISA, vip Kit ELISA, vip.L Kit ELISA
- Sujet
- Vasoactive intestinal peptide (VIP) is a peptide hormone containing 28 amino acid residues and is produced in many areas of the human body including the gut, pancreas and suprachiasmatic nuclei of the hypothalamus in the brain.It has a half-life in the blood of about two minutes. With respect to the digestive system, VIP seems to induce smooth muscle relaxation (lower esophageal sphincter, stomach, gallbladder), stimulate secretion of water into pancreatic juice and bile, and cause inhibition of gastric acid secretion and absorption from the intestinal lumen. Its role in the intestine is to greatly stimulate secretion of water and electrolytes, as well as dilating intestinal smooth muscle, dilating peripheral blood vessels, stimulating pancreatic bicarbonate secretion, and inhibiting gastrin-stimulated gastric acid secretion. These effects work together to increase motility.It also has the function of stimulating pepsinogen secretion by chief cells. It is also found in the brain and some autonomic nerves. One region of the brain includes a specific area of the suprachiasmatic nuclei (SCN), the location of the 'master circadian pacemaker'. The SCN coordinates daily timekeeping in the body and VIP plays a key role in communication between individual brain cells within this region. Further, VIP is also involved in synchronising the timing of SCN function with the environmental light-dark cycle. Combined, these roles in the SCN make VIP a crucial component of the mammalian circadian timekeeping machinery. VIP helps to regulate prolactin secretion.It is also found in the heart and has significant effects on the cardiovascular system. It causes coronary vasodilation as well as having a positive inotropic and chronotropic effect. Research is being performed to see if it may have a beneficial role in the treatment of heart failure.
- ID gène
- 3180
- Pathways
- Hormone Activity, cAMP Metabolic Process
-